Beckman Coulter Mishima K.K. Recalls
2 recalls on record in the United States
Beckman Coulter Mishima K.K. recalls CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B232
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment
View recall details →
Beckman Coulter Mishima K.K. recalls DxC 700 AU, REF: B86444, B86446
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.
View recall details →