Becton, Dickinson And Company Recalls
3 recalls on record in the United States
Becton, Dickinson and Company recalls Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
View recall details →
Becton, Dickinson and Company recalls Swan-Ganz Pacing Catheter, Models: D200F7;
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
View recall details →
Becton, Dickinson and Company recalls Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
View recall details →