DATEX--OHMEDA Recalls
7 recalls on record in the United States
DATEX--OHMEDA, INC. recalls GE Healthcare Giraffe Incubator Carestation CS1, Infant Incu
Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.
View recall details →
DATEX--OHMEDA, INC. recalls GE Healthcare Giraffe Omnibed Carestation CS1, combination i
Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.
View recall details →
DATEX--OHMEDA, INC. recalls GE Heathcare Giraffe OmniBed; neo natal incubator
GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.
View recall details →
DATEX--OHMEDA, INC. recalls GE Heathcare Giraffe OmniBed Carestation, Model Number 20828
GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.
View recall details →
DATEX--OHMEDA, INC. recalls Giraffe OmniBed Carestation CS1, Model Numbers: a. 208284
GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices
View recall details →
DATEX--OHMEDA, INC. recalls Beside panel FRU (Field Replacement Units) Kits
GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices
View recall details →
DATEX--OHMEDA, INC. recalls Giraffe OmniBed, with the following Model Numbers: 6650-0004
GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices
View recall details →