2026-03-25
Diagnostica Stago, Inc. recalls Brand Name: STA Liatest D-Di Product Name: STA Liatest D-Di
After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
View recall details →
medical
Class II
2026-09-09
Diagnostica Stago, Inc. recalls STA Staclot Protein S, Model/Catalog Number: 00746-Reagent L
A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified