GE Healthcare LLC Recalls
7 recalls on record in the United States
GE Healthcare LLC recalls SIGNA Premier systems
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
View recall details →
GE Healthcare recalls Centricity PACS Server V7, intended for managing digital med
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
View recall details →
GE Healthcare recalls Centricity PACS V2, V3, V4, V6, V7, software product is inte
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
View recall details →
GE Healthcare recalls Centricity Enterprise Archive V2, V3, and V4, software produ
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
View recall details →
GE Healthcare recalls Core Services Hub V8, a set of backend services including Cr
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
View recall details →
GE Healthcare recalls Workflow Core Services V8, intended to provide set of servic
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
View recall details →
GE Healthcare recalls Enterprise Archive V8, software product for receiving, archi
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
View recall details →