Hamilton Medical AG Recalls
2 recalls on record in the United States
Hamilton Medical AG recalls Breathing circuit set, coaxial: REF: 260127, 260128, 260167,
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement
View recall details →
Hamilton Medical AG recalls IntelliCuff, Portable automatic cuff pressure controller RE
Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.