Heraeus Medical GmbH (Dental Division) Recalls
2 recalls on record in the United States
Heraeus Medical GmbH (Dental Division) recalls Heraeus, PALAMIX uno. Material Number: 66057893.
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
View recall details →
Heraeus Medical GmbH (Dental Division) recalls Heraeus, PALAMIX duo. Material Number: 66057897.
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).