Ion Beam Applications S.A. Recalls

1 recall on record in the United States

medical: 1
medical Class II
2026-03-18

Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235

It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.

View recall details →