Magellan Diagnostics, Inc. Recalls

3 recalls on record in the United States

medical: 3
medical Class II
2026-04-29

Magellan Diagnostics, Inc. recalls LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004

Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.

View recall details →

medical Class II
2026-04-29

Magellan Diagnostics, Inc. recalls (1)(2) LeadCare¿ II Blood Lead Test Kit, (3)LeadCare¿ II 5X

Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.

View recall details →

medical Class II
2026-04-29

Magellan Diagnostics, Inc. recalls LeadCare Ultra Blood Lead Test Kit; Catalog Number: 70-8098

Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.

View recall details →