Restor3d Recalls
7 recalls on record in the United States
restor3d Inc. recalls Evans Wedge Disposable Trials, Medium Footprint, 10mm Thickn
During re-work of disposable instruments to relabel for 4-year shelf-life of disposable instruments, parts got mixed into other lots.
View recall details →
restor3d Inc. recalls Non-Sterile Double Ended Trial Kit, Small Footprint, (6110-N
Product labeled incorrectly. The lot was labeled with Small footprint when the product was Medium.
View recall details →
restor3d Inc. recalls Evans Wedge Disposable Trials, Large Footprint, 8mm Thicknes
During re-work of disposable instruments to relabel for 4-year shelf-life of disposable instruments, parts got mixed into other lots.
View recall details →
restor3d Inc. recalls Brand Name: Veritas Standard Reverse Total Shoulder Product
Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.
View recall details →
restor3d Inc. recalls Kinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut
Ankle implantation systems were packaged with incorrect patient-specific cutting guides, which may result in: delayed surgery, the need to use standard instrumentation, reduced fixation or poor alignment of implants, implant loosening/migration, or need for revision surgery.
View recall details →
restor3d Inc. recalls Generic Name: restor3d MTP Implant Generic Name: MTP Implan
Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.
View recall details →
restor3d Inc. recalls Veritas Reverse Total Shoulder Replacement, Model/Catalog N
(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
View recall details →