Spacelabs Healthcare, Inc. Recalls

1 recall on record in the United States

medical: 1
medical Class II
2026-04-15

Spacelabs Healthcare, Inc. recalls Multi-parameter Command Module, Model 91496, optioned with M

Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.

View recall details →