medical Class II Updated 2024-10-09

Ossur H / F recalls Miami J Select Collar, Catalog: MJS-101; Miami J Select Coll

Recalled Product

Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101


Hazard / Issue

Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.

Issued by

FDA

Affected States: NH, MA, NY, NJ, PA, GA, MD, NC, IN, FL, AR, TN, KY, OH, MI, IL, TX, MO, CO, UT, AZ, WA, OR, CA, NM, VA, IA, MN, NE, SC, MT, LA, ID, SD, AL, CT, OK, AK
Lot/Code Info: REF/UDI-DI/GTIN/Lot Range/Sold in the Period, Date Range: MJS-101/05690967817008/MX220516 to MX230104/2022-05-19 to 2023-06-05, MJSR-101/05690967817145/MX220516 to MX230104/2022-05-19 to 2023-06-05