medical Class II Updated 2024-10-09

Philips North America recalls IntelliVue G7m Anesthesia Gas Module, Product Number 866173

Recalled Product

IntelliVue G7m Anesthesia Gas Module, Product Number 866173


Hazard / Issue

The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00884838051492; Serial Numbers: DE43800129 through DE43812627
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →