medical Class II Updated 2018-10-10

Radiometer America Inc recalls ABL90 FLEX analyzers with software versions 3.3MR1 or below.

Recalled Product

ABL90 FLEX analyzers with software versions 3.3MR1 or below. The ABL90 FLEX Analyzer is a portable, automated Analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood. It is intended for use in a laboratory environment, near patient or point-of-care setting.


Hazard / Issue

The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: ¿ The analyzer is in ready mode ¿ Operator lifts the inlet and then closes it again, thereby initiating a rinse ¿ Immediately after lifts the inlet and quickly closes it again

Issued by

FDA

Distribution: US Distribution and Internationally to Canada.
Lot/Code Info: ABL90 FLEX analyzers with software versions 3.3MR1 or below.
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →