Radiometer America Inc recalls ABL90 FLEX analyzers with software versions 3.3MR1 or below.
Recalled Product
ABL90 FLEX analyzers with software versions 3.3MR1 or below. The ABL90 FLEX Analyzer is a portable, automated Analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood. It is intended for use in a laboratory environment, near patient or point-of-care setting.
Hazard / Issue
The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: ¿ The analyzer is in ready mode ¿ Operator lifts the inlet and then closes it again, thereby initiating a rinse ¿ Immediately after lifts the inlet and quickly closes it again
Issued by
FDA
Were you affected by this recall?
If you were harmed by a recalled product, you may have legal options. Consider speaking with a consumer protection lawyer in your state.
Looking for safe alternatives?
Browse certified-safe products in this category on Amazon →
Get the next US recall alert before it hits the news.
Free. Weekly digest or instant alerts. Unsubscribe anytime.