Smiths Medical ASD Inc. recalls CADD Medication Cassette Reservoir, 100mL, Product Code 21-7
Recalled Product
CADD Medication Cassette Reservoir, 100mL, Product Code 21-7002-24
Hazard / Issue
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Issued by
FDA
Distribution: Worldwide distribution.
Lot/Code Info: UDI/DI 1061058602228, Lot Numbers: 3977441, 4013370, 3983314, 4013371, 4013369, 4013372, 4025948, 4033993, 4037444, 4040238