medical Class II Updated 2024-10-30

Smiths Medical ASD Inc. recalls CADD Medication Cassette Reservoir, 100mL, FS, YELLOW, Produ

Recalled Product

CADD Medication Cassette Reservoir, 100mL, FS, YELLOW, Product Code 21-7300-24


Hazard / Issue

Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.

Issued by

FDA

Distribution: Worldwide distribution.
Lot/Code Info: UDI/DI 10610586027192, Lot Numbers: 3971732, 3971731, 4034016, 3977457, 4013365, 4019421, 4019422, 4034015, 4037744, 4037745, 4034018, 4034017, 4046844, 4046843, 4053912, 4048899
View official government recall