medical Class II Updated 2024-10-30

Smiths Medical ASD Inc. recalls CADD Medication Cassette Reservoir, 50mL, Product Code 21-70

Recalled Product

CADD Medication Cassette Reservoir, 50mL, Product Code 21-7001-24


Hazard / Issue

Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.

Issued by

FDA

Distribution: Worldwide distribution.
Lot/Code Info: UDI/DI 10610586022111, Lot Numbers: 4040237, 4013368, 3983342, 4046817, 405240, 4060903