medical Class II Updated 2018-10-10

Leica Microsystems, Inc. recalls Leica HistoCore SPECTRA CV Coverslipper, used in preparation

Recalled Product

Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microscopic analysis for medical diagnostics.


Hazard / Issue

Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock.

Issued by

FDA

Affected States: MO
Lot/Code Info: Serial Number 102
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →