medical HIGH RISK Updated 2025-10-15

NOXBOX LTD recalls NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX

See all recalls from NOXBOX LTD →

Recalled Product

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

Issued by

FDA

Affected States: TN
Lot/Code Info: Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.