medical Class II Updated 2021-10-13

Beckman Coulter Inc. recalls Access hsTnI High Sensitivity Troponin I, REF B52699 Acce

Recalled Product

Access hsTnI High Sensitivity Troponin I, REF B52699 Access Immunoassay System REAGENT PACK


Hazard / Issue

Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover predominantly into the particle well of the reagent pack by way of the probe. The particle well then becomes contaminated by the high sample from the probe. The extent of the carryover is dependent on sample concentration and frequency of high concentration samples.

Issued by

FDA

Affected States: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, NC, ND, NE, NV, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, WV, WY
Lot/Code Info: Catalog Number:B52699 Lot Number / UDI Code: 921491 / (01)15099590693183(17)201130(11)191201(10)921491; 921638 / (01)15099590693183(17)201231(11)200101(10)921638; 921631 / (01)15099590693183(17)210131(11)200201(10)921631; 921750 / (01)15099590693183(17)210131(11)200201(10)921750; 921805 / (01)15099590693183(17)210228(11)200229(10)921805; 921831 / (01)15099590693183(17)210331(11)200331(10)921831; 921943 / (01)15099590693183(17)210430(11)200430(10)921943; 922000 / (01)15099590693183(17)210531(11)200531(10)922000; 922039 / (01)15099590693183(17)210630(11)200630(10)922039; 922040 / (01)15099590693183(17)210731(11)200731(10)922040; 922062 / (01)15099590693183(17)210731(11)200731(10)922062; 922145 / (01)15099590693183(17)210831(11)200831(10)922145; 922146 / (01)15099590693183(17)210930(11)200930(10)922146; 922338 / (01)15099590693183(17)210930(11)200930(10)922338; 922440 / (01)15099590693183(17)211031(11)201031(10)922440; 922545 / (01)15099590693183(17)211031(11)201031(10)922545; 922547 / (01)15099590693183(17)211130(11)201130(10)922547; 922713 / (01)15099590693183(17)211231(11)201231(10)922713; 922714 / (01)15099590693183(17)211231(11)201231(10)922714; 922722 / (01)15099590693183(17)220131(11)210131(10)922722; 922897 / (01)15099590693183(17)220131(11)210131(10)922897; 124235 / (01)15099590693183(17)220228(11)210228(10)124235; 124339 / (01)15099590693183(17)220228(11)210228(10)124339; 922898 / (01)15099590693183(17)220228(11)210228(10)922898; 124340 / (01)15099590693183(17)220331(11)210331(10)124340; 124348 / (01)15099590693183(17)220430(11)210430(10)124348; 124353 / (01)15099590693183(17)220430(11)210430(10)124353; 124688 / (01)15099590693183(17)220430(11)210430(10)124688; 124354 / (01)15099590693183(17)220531(11)210531(10)124354; 124355 / (01)15099590693183(17)220630(11)210630(10)124355; 124356 / (01)15099590693183(17)220630(11)210630(10)124356; 124357 / (01)15099590693183(17)220630(11)210630(10)124357;
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →