medical Class II Updated 2023-10-11

AB Sciex, LLC recalls API3200MD Mass Spectrometer, Part Number 4466230

Recalled Product

API3200MD Mass Spectrometer, Part Number 4466230


Hazard / Issue

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: EQ230011907, EQ230021911, EQ230001901
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →