medical Class II Updated 2023-10-11

AB Sciex, LLC recalls Citrine QTRAP Mass Spectrometer, Part Number 5063685

Recalled Product

Citrine QTRAP Mass Spectrometer, Part Number 5063685


Hazard / Issue

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00842365100540; Serial Numbers: DD250051901, DD250172009, DD250182009, DD230001804, DD250111910, DD250132004, DD250091909, DD250101910, DD250192010, CN10011707PT
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →