medical Class II Updated 2023-10-11

AB Sciex, LLC recalls Citrine Triple Quad Mass Spectrometer, Part Number 5063684

Recalled Product

Citrine Triple Quad Mass Spectrometer, Part Number 5063684


Hazard / Issue

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →