medical Class II Updated 2020-10-14

Siemens Healthcare Diagnostics, Inc. recalls Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reage

Recalled Product

Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in human plasma Catalog Number: RF627, SMN# 10471068


Hazard / Issue

Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin compared with a control lot was +25%. A maximum bias of +34% in patient samples around the 99th percentile observed

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot # EB0255 UDI: 00842768037726EB025520091110471068840
View official government recall