medical Class II Updated 2018-10-10

Ortho Clinical Diagnostics Inc recalls VITROS Immunodiagnostic Products - FSH Reagent Pack (107587

Recalled Product

VITROS Immunodiagnostic Products - FSH Reagent Pack (10758750000302) 1931922


Hazard / Issue

The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 2470 2480 2490 2500 2510 2520 2530
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →