medical Class II Updated 2018-10-10

Ortho Clinical Diagnostics Inc recalls VITROS Immunodiagnostic Products - TSH Reagent Pack (107587

Recalled Product

VITROS Immunodiagnostic Products - TSH Reagent Pack (10758750000227) 1912997


Hazard / Issue

The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 5535 5540 5550 5560 5570 5580 5590 5600 5610 5615 5620 5630 5640 5650 5655 5660 5665 5670 5675 5678 5680 5685 5690
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →