medical Class II Updated 2019-10-16

Becton, Dickinson and Company, BD Biosciences recalls BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF

Recalled Product

BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 662877, when using BD Trucount Tubes


Hazard / Issue

Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.

Issued by

FDA

Affected States: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE
Lot/Code Info: BD FACSLyric Flow Cytometer serial numbers: Z662877000001, Z662877000002, Z662877000003, Z662877000004, Z662877000005, Z662877000006, Z662877000007, Z662877000008.