medical MODERATE Updated 2026-10-14

Steris Corporation recalls VERIFY Dual Species Self-Contained Biological Indicator Mod

See all recalls from Steris Corporation →

Recalled Product

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI: 50724995023824 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.