Biomerieux Inc recalls VITEK2 GP-ID, REF 21342 IVD

See all recalls from Biomerieux Inc →

Recalled Product

VITEK2 GP-ID, REF 21342 IVD

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Customers have reported, and an internal investigation has confirmed, occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC)R) 700327(TM) in association with three (3) VITEK(R) 2 GP ID Lots, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. In accordance with the VITEK(R) 2 GP ID Instructions For Use, the expected O129R reaction for ATCC(R) 700327(TM) strain is positive. As a precaution, bioM¿rieux is also including in the scope of this notice all GP ID card lots manufactured with the same raw materials.

Issued by

FDA

Distribution: worldwide
Lot/Code Info: UDI 03573026131920, Lot Number 2420348103