medical MODERATE Updated 2026-10-14

Steris Corporation recalls VERIFY Biological Steam Test Pack Model/Catalog Number: S30

See all recalls from Steris Corporation →

Recalled Product

VERIFY Biological Steam Test Pack Model/Catalog Number: S3065

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI: 10724995220300 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for VERIFY Biological Steam Test →
View official government recall