medical Class II Updated 2024-10-23

LEICA BIOSYSTEMS NUSSLOCH GMBH recalls Leica Biosystems Cryostat Model Number CM1900, Material Numb

Recalled Product

Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2; for freezing and sectioning tissue samples


Hazard / Issue

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: a) Material Number 14045227382, UDI/DI 04049188075840, Serial Numbers: 1198121999, 383051996; b) Material Number 14901900100, UDI/DI 04049188003324, Serial Numbers: 703111997, 1018021999, 1538042001, 2058, 629071997, 2994, 1170111999, 3686102007, 3698; c) Material Number 14901900700, UDI/DI 04049188003300, Serial Numbers: 1267032000, 2602062004, 1577052001, 1679092001, 2542042004, 1771, 1885032002, 2682092004, 2376112003, 2774122004, 1114, 1749112001, 1997072002; d) Material Number 1491900UVU2, UDI/DI 04049188002570, Serial Numbers: 3243052006
View official government recall