medical Class II Updated 2022-10-26

B. Braun Medical, Inc. recalls CARESITE SMALLBORE EXT SET T-PORT 5 IN., Product Code 470114

Recalled Product

CARESITE SMALLBORE EXT SET T-PORT 5 IN., Product Code 470114, 100 units per carton. The A6 Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood.


Hazard / Issue

The extension set has a high probability of potential failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI (GTIN): 04046964298691 Lots/Expiration Dates 61762215 15-Dec-23 61762791 22-Jan-24 61767369 26-Feb-24 61779625 7-Jun-24 61791895 11-Aug-24 61806047 4-Nov-24
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →