Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance System, AV

Recalled Product

Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EVAL US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which includes intravascular ultrasound (IVUS). When used with a peripheral vascular imaging catheter, the system utilizes the IVUS modality for ultrasound examination of intravascular pathology.


Hazard / Issue

Boston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Material Number H7492493120D0, GTIN/UDI/DI 00191506033149, Batch/Serial Numbers: 102899658, 103225132, 103246673, 103647916, 104043675, 104072168, 104258818, 104512059
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →