medical Class II Updated 2022-10-26

Medicina Uk Ltd recalls Medicina Reusable Oral syringe 1ml Code: OTH01

Recalled Product

Medicina Reusable Oral syringe 1ml Code: OTH01


Hazard / Issue

Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: (01)05056115402869 (01)05056115402869(17)260700(10)HM21053 Batch number: HM18041, HM21053