medical Class II Updated 2024-10-23

CHANGE HEALTHCARE CANADA COMPANY recalls Change Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 1

Recalled Product

Change Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.


Hazard / Issue

A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI Numbers: (01)17540262100013(10)140100, (01)17540262100020(10)140101, (01)17540262100037(10)140200, (01)17540262100044(10)140300, (01)17540262100051(10)140302, (01)17540262100099(10)150000, (01)17540262100105(10)150001, (01)17540262100129(10)150100. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →