medical Class II Updated 2020-10-28

Siemens Healthcare Diagnostics, Inc. recalls Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex rea

Recalled Product

Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro diagnostic test for the quantitative measurement of alkaline phosphatase in human serum and plasma Catalog Number: K2115, SMN# 10642444


Hazard / Issue

Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was approximately -27% at a concentration of 129 U/L (2.15 ukat/L). The low outlier results are generated from the last five (5) tests of the well

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number/UDI/Exp. Date: 1) 19247AB UDI: 0084276803903419247AB20090310642444840 Exp.2020-09-03, (2) 19282BB UDI: 0084276803903419282BB20100810642444840 Exp. 2020-10-08, (3) 19330BD UDI:0084276803903419330BD20112510642444840 Exp. 2020-11-25. Expanded Recall: 3/10/21 20337AB, 20311BA, 20275AA, 20218AA, 20203BA, 20170AA, 20125BA, 20078AB.
View official government recall

Looking for safe alternatives?