medical Class II Updated 2024-10-30

SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalls Trinias unity, Digital Angiography System with Catheterizati

Recalled Product

Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component.


Hazard / Issue

If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 04540217058327, 04540217049066, 04540217058143 Serial Numbers: 41EFCD7AC001, 41E58C773001, 41EFCD79C001, 41EFCD78B001, 41EFC73A9001, 41EFCD7D4001, 41E58C76B001, 41EFC73D9001, 41EFCD77C001, 41EFC73A5001, 41EFC73C5001, 41EFCD79A001, 41EFCD7A6001.