medical MODERATE Updated 2026-10-21

SI-BONE, Inc. recalls iFuse TORQ Implant System Model/Catalog Number: 10055T(10x5

See all recalls from SI-BONE, Inc. →

Recalled Product

iFuse TORQ Implant System Model/Catalog Number: 10055T(10x55mm), 10060T(10x60mm), 10065T(10x65mm), 10080T(10x55mm), 11580T(11.5x80mm)

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 10055T-UDI:00810055520077/Lot:9142871 10060T-UDI:00810055520084/Lot:9137761 10065T-UDI:00810055520091/Lot:9139213 10080T-UDI:00810055520121/Lot:9137781 11580T-UDI:00810055520367/Lot:9142911

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.