medical Class II Updated 2020-10-28

Smith & Nephew, Inc. recalls RAPTORMITE 3.7MM AB W/ NEEDLES AND TWO-Absorbable Suture anc

Recalled Product

RAPTORMITE 3.7MM AB W/ NEEDLES AND TWO-Absorbable Suture anchor Product Number: 72201805


Hazard / Issue

Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by the pouch

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Batch Numbers: 2025354 2031116 2040066 2040702 2042557 2049638 2053749 2055239