medical Class II Updated 2023-11-01

Omega Medical Imaging, Inc. recalls Omega Medical Imaging Elevating Monitor Suspension Systems,

Recalled Product

Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085


Hazard / Issue

An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower boom became detached at the point shown and was restrained by the installed safety cable.

Issued by

FDA

Distribution: Alabama, California, Illinois, Indiana, Kentucky, Louisiana, Missouri, Montana, North Carolina, Ohio, Oklahoma
Lot/Code Info: all units manufactured between 10-22-2012 and 07-09-2019