medical MODERATE Updated 2026-10-21

Beckman Coulter, Inc. recalls DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513

See all recalls from Beckman Coulter, Inc. →

Recalled Product

DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Product/UDI-DI: DxU 840m/15099590748654, DxU 850m/15099590748647. All Serial Numbers

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for DxU 840m Iris, REF: →
View official government recall