medical Class II Updated 2018-10-24

Mindray DS USA, Inc. dba Mindray North America recalls Mindray BeneVision Distributed Monitoring System (DMS), Part

Recalled Product

Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00


Hazard / Issue

The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only 32 transmitters at a time. For the issue to manifest, the BeneVision DMS needs to be configured to use flexible device assignment, have more than 64 TD60 transmitters, and the workflow must allow 64 TD60s to move around between multiple central stations in the same care area.

Issued by

FDA

Affected States: CA, CO, FL, MA, NY, PA, TX, WV
Lot/Code Info: This potential issue affects BeneVision DMS with software version 03.01.00 to 03.03.00