medical MODERATE Updated 2022-11-09

MICROVENTION INC. recalls WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Compon

See all recalls from MICROVENTION INC. →

Recalled Product

WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: REF/UDI/Lots: WDC-2/00842429103128/0000095683, 0000096284, 0000205967, 0000206629, 0000215962, 0000216862, 0000224074, 0000241662, 0000242343, 0000242345; Distributed in China: WDC/00842429113684/0000078840, 0000086571, 0000088347, 0000088349, 0000136477, 0000181460, 0000191147; Distribution in Japan: MV-WDC2/04987892122231/0000142172; 0000166445; 0000210109; 0000245337

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
View official government recall