medical Class II Updated 2021-11-03

Xstrahl Limited recalls Concerto User Interface Software provided with the following

Recalled Product

Concerto User Interface Software provided with the following systems: (1) Xstrahl 100 Electronic Brachytherapy - 100kV Superficial X-Ray Therapy System; (2) Xstrahl 150 Electronic Brachytherapy - 150kV Superficial X-Ray Therapy System; (3) Xstrahl 200 - 225kV Superficial / Orthovoltage X-Ray Therapy System; (4) Xstrahl 300 - 300kV Orthovoltage X-Ray Therapy System; (5) Xstrahl X 80 Photoelectric Therapy System.


Hazard / Issue

If a saved treatment plan with 2 opposing beams is edited prior to approval, then Beam 2 is not updated with the changed parameters upon selecting save, resulting in error messages during the treatment and possible mis-treatment.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: V2.0, V2.1, and V2.2 when 2 opposing beam treatment plans are used.
View official government recall