medical Class II Updated 2021-11-03

AB SCIEX recalls Cliquid MD version 3.4 software used in conjunction with Ana

Recalled Product

Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem mass spectrometry (LC-MS/MS) systems. Model No. 5088288


Hazard / Issue

The values of the Internal Standard (IS) concentrations are incorrectly derived when the user builds a customized test, then set the IS column to hidden and left the column empty. This may lead to the user making incorrect conclusions using incorrect results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Software version 3.4