medical Class II Updated 2024-10-30

MEDLINE INDUSTRIES, LP - Northfield recalls Medline brand medical procedure convenience kits, labeled as

Recalled Product

Medline brand medical procedure convenience kits, labeled as: a) LITHOTOMY PACK, REF DYNJ24362C; b) MAJOR GYNE LAPAROSCOPY SHSC, REF DYNJ46774B; c) GYN ANCILARY PACK, REF DYNJ84509; d) GYN LITHOTOMY MINOR, REF DYNJ907707


Hazard / Issue

It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: a) REF DYNJ24362C, UDI/DI 40889942830054 (each), 10889942830053 (case), Lot Numbers: 24ABG207; b) REF DYNJ46774B, UDI/DI 40889942582748 (each), 10889942582747 (case), Lot Numbers: 23LBN531; c) REF DYNJ84509, UDI/DI 40195327372881 (each), 10195327372880 (case), Lot Numbers: 23LBL712; d) REF DYNJ907707, UDI/DI 40193489848978 (each), 10193489848977 (case), Lot Numbers: 24ABR224
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →