medical Class II Updated 2019-10-30

Life Technologies Holdings Pte Ltd recalls Torrent Suite Dx Software Version (IUO) 5.4 Catalog Number:

Recalled Product

Torrent Suite Dx Software Version (IUO) 5.4 Catalog Number: A31774 OEM Customer Software: Sentosa SQ Suite Software v5.6.15


Hazard / Issue

Torrent SuiteTM Dx analysis software, a component of the Ion PGM Dx Instrument System, incorrectly displays a positive result when a different analytical or de novo variant is detected at the same locus as a specified clinical variant

Issued by

FDA

Affected States: AZ, CA, CO, CT, FL, GA, IN, MA, MD, NC, NJ, NM, NY, OR, TX, VA, WA
Lot/Code Info: The field action is associated with the software available on the PGM Dx instrument, therefore no lot code is applicable for this part type