medical Class II Updated 2020-10-28

ICU Medical, Inc. recalls 7" (18cm) Appx 0.31 ml, Smallbore Ext Set w/Remv MicroClave¿

Recalled Product

7" (18cm) Appx 0.31 ml, Smallbore Ext Set w/Remv MicroClave¿ Clear, NanoClave T-Connector, Clamp, Rotating Luer. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709051055(17)250101(30)50(10)4590834; (01)10887709051055(17)250101(30)50(10)4557481; (01)10887709051055(17)241201(30)50(10)4536601; (01)10887709051055(17)250201(30)50(10)4724818; (01)10887709051055(17)250201(30)50(10)4733548. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).


Hazard / Issue

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers:4590834, 4557481, 4536601, 4724818, 4733548.
View official government recall