medical Class II Updated 2020-10-28

ICU Medical, Inc. recalls 10" (25 cm) Appx 2.8 ml, Trifuse Ext Set w/3 MicroClave¿ Cle

Recalled Product

10" (25 cm) Appx 2.8 ml, Trifuse Ext Set w/3 MicroClave¿ Clear, Tri-Connector, NanoClave¿ T-Connector, 4 Clamps (Red, White, Blue), Rotating Luer;1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709052816(17)241201(30)50(10)4537141 The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).


Hazard / Issue

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number: 4537141.