medical Class I Updated 2023-11-15

Olympus Corporation of the Americas recalls BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF

Recalled Product

BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF-1T60, BF-P150 BF-P60 and EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number XT160


Hazard / Issue

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model BF-1T150 UDI-DI: 4953170308185; Model BF-1T60 UDI-DI: 4953170339264; Model BF-P150 UDI-DI: N/A; Model BF-P60 UDI-DI: 4953170339196; Model BF-XT160 UDI-DI: N/A
View official government recall